-10-SENSOR REPLACEMENTSTEP 4)Remove sensor from sensor cup and remove leads . Takethe new sensor and remove the metal spring. Reattachleads to the pr
-11-CALIBRATION PROCEDUREDo not use inert gases to zero the monitor. This will cause premature failure of the sensor.CO Zero AdjustmentTo zero the i
-12-CO SPAN ADJUSTMENTUse only 10 - 20 ppm CO gas for calibration. Using a higher concentration may decrease accuracy atlower scale readings. Note:
-13-MODEL BB15-CO REPLACEMENT FILTER BREAKDOWNNOTE: FOR UNITS PURCHASED PRIOR TO 12/04, PLEASE CONTACTCUSTOMER SERVICE FOR CORRECT ORDERING INFORMATI
-14-MODEL BB30-CO REPLACEMENT FILTER BREAKDOWNITEM # DESCRIPTION P/N1 FILTER CHANGE INDICATOR WL2612 FILTER BOWL O-RING WL2663 “A” FILTER ELEMENT BB30
-15-MODEL BB50-CO REPLACEMENT FILTER BREAKDOWNITEM # DESCRIPTION P/N1 FILTER CHANGE INDICATOR WL0562 FILTER BOWL O-RING WL0913 “A” FILTER ELEMENT BB50
-16-MODEL BB75-CO REPLACEMENT FILTER BREAKDOWNITEM # DESCRIPTION P/N1 FILTER CHANGE INDICATOR WL2612 “A” FILTER ELEMENT AND O-RING BB75-A3 “C” FILTER
-17-MODEL BB100-CO REPLACEMENT FILTER BREAKDOWNITEM # DESCRIPTION P/N1 FILTER CHANGE INDICATOR WL0562 FILTER BOWL O-RING WL1133 “A” FILTER ELEMENT BB1
-18-MODEL BB150-CO REPLACEMENT FILTER BREAKDOWNITEM # DESCRIPTION P/N1 FILTER CHANGE INDICATOR WL0562 FILTER BOWL O-RING WL1133 “A” FILTER ELEMENT BB1
-19-Warranty DisclaimerAir Systems’ manufactured equipment is warranted to the original user against defects in work-manship or materials under normal
-2-*Note: Some models may not have respirator connections. They can be ordered with NPT outlets for connectionto drop stations or point of attachmen
-3-BREATHER BOXTM PARTS IDENTIFICATIONMODELS BB30-100MODEL BB15-COMODEL BB150-CO
-4-BREATHING AIR QUALITY POSITION STATEMENTThe responsibility for the quality of breathing air rests with the user. Compliance with federal, state,
-5-Test CircuitManually activatedSensor TypeSealed electrochemical sensor for Carbon MonoxideAccuracy +/- 1% full scaleResponse90% in 10-15 secondsDet
-6-ITEM # DESCRIPTION PART #1 LCD DISPLAY MONC7032 SPAN POTENTIOMETER MONC702A3 ALARM SET POINT POTENTIOMETER MONC702A4 ZERO POTENTIOMETER MONC7025 AI
-7-BREATHER BOXTMSETUP AND OPERATIONSTEP 1)Secure a primary air source of sufficient air flow and discharge pressure.The number and type of respirator
-8-STEP 6)Connect the air source to the inlet fitting.STEP 7)Hold the “ON/OFF/TEST” switch in the “TEST” position. All local andremote audible/visual
-9-SHUTDOWN1) Make sure all personnel have egressed from the work area.2) Shut off air source to the box.3) Remove air pressure from the box by pullin